Costanzo, Alberto

Link to this page

Authority KeyName Variants
d0132d0c-b74a-4764-8cc0-64e601f13fee
  • Costanzo, Alberto (1)
Projects
No records found.

Author's Bibliography

Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments

Plotino, Gianluca; Costanzo, Alberto; Grande, Nicola M.; Petrović, Renata; Testarelli, Luca; Gambarini, Gianluca

(Elsevier Science Inc, New York, 2012)

TY  - JOUR
AU  - Plotino, Gianluca
AU  - Costanzo, Alberto
AU  - Grande, Nicola M.
AU  - Petrović, Renata
AU  - Testarelli, Luca
AU  - Gambarini, Gianluca
PY  - 2012
UR  - https://smile.stomf.bg.ac.rs/handle/123456789/1753
AB  - Introduction: The purpose of this study was to evaluate the effect of autoclave sterilization on cyclic fatigue resistance of rotary endodontic instruments made of traditional and new nickel-titanium (NiTi) alloys. Methods: Four NiTi rotary endodontic instruments of the same size (tip diameter 0.40 mm and constant .04 taper) were selected: K3, Mtwo, Vortex, and K3 XF prototypes. Each group was then divided into 2 subgroups, unsterilized instruments and sterilized instruments. The sterilized instruments were subjected to 10 cycles of autoclave sterilization. Twelve files from each different subgroup were tested for cyclic fatigue resistance. Means and standard deviations of number of cycles to failure (NCF) and fragment length of the fractured tip were calculated for each group, and data were statistically analyzed (P  lt  .05). Results: Comparing the results between unsterilized and sterilized instruments for each type of file, differences were statistically significant (P  lt  .05) only between sterilized and unsterilized K3XF files (762 versus 651 NCF). The other instruments did not show significant differences (P > .05) in the mean NCF as a result of sterilization cycles (K3, 424 versus 439 NCF; Mtwo, 409 versus 419 NCF; Vortex, 454 versus 480 NCF). Comparing the results among the different groups, K3 XF (either sterilized or not) showed a mean NCF significantly higher than all other files(P  lt  .05). Conclusions: Repeated cycles of autoclave sterilization do not seem to influence the mechanical properties of NiTi endodontic instruments except for the K3 XF prototypes of rotary instruments that demonstrated a significant increase of cyclic fatigue resistance. (J Endod 2012;38:222-225)
PB  - Elsevier Science Inc, New York
T2  - Journal of Endodontics
T1  - Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments
VL  - 38
IS  - 2
SP  - 222
EP  - 225
DO  - 10.1016/j.joen.2011.10.017
ER  - 
@article{
author = "Plotino, Gianluca and Costanzo, Alberto and Grande, Nicola M. and Petrović, Renata and Testarelli, Luca and Gambarini, Gianluca",
year = "2012",
abstract = "Introduction: The purpose of this study was to evaluate the effect of autoclave sterilization on cyclic fatigue resistance of rotary endodontic instruments made of traditional and new nickel-titanium (NiTi) alloys. Methods: Four NiTi rotary endodontic instruments of the same size (tip diameter 0.40 mm and constant .04 taper) were selected: K3, Mtwo, Vortex, and K3 XF prototypes. Each group was then divided into 2 subgroups, unsterilized instruments and sterilized instruments. The sterilized instruments were subjected to 10 cycles of autoclave sterilization. Twelve files from each different subgroup were tested for cyclic fatigue resistance. Means and standard deviations of number of cycles to failure (NCF) and fragment length of the fractured tip were calculated for each group, and data were statistically analyzed (P  lt  .05). Results: Comparing the results between unsterilized and sterilized instruments for each type of file, differences were statistically significant (P  lt  .05) only between sterilized and unsterilized K3XF files (762 versus 651 NCF). The other instruments did not show significant differences (P > .05) in the mean NCF as a result of sterilization cycles (K3, 424 versus 439 NCF; Mtwo, 409 versus 419 NCF; Vortex, 454 versus 480 NCF). Comparing the results among the different groups, K3 XF (either sterilized or not) showed a mean NCF significantly higher than all other files(P  lt  .05). Conclusions: Repeated cycles of autoclave sterilization do not seem to influence the mechanical properties of NiTi endodontic instruments except for the K3 XF prototypes of rotary instruments that demonstrated a significant increase of cyclic fatigue resistance. (J Endod 2012;38:222-225)",
publisher = "Elsevier Science Inc, New York",
journal = "Journal of Endodontics",
title = "Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments",
volume = "38",
number = "2",
pages = "222-225",
doi = "10.1016/j.joen.2011.10.017"
}
Plotino, G., Costanzo, A., Grande, N. M., Petrović, R., Testarelli, L.,& Gambarini, G.. (2012). Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments. in Journal of Endodontics
Elsevier Science Inc, New York., 38(2), 222-225.
https://doi.org/10.1016/j.joen.2011.10.017
Plotino G, Costanzo A, Grande NM, Petrović R, Testarelli L, Gambarini G. Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments. in Journal of Endodontics. 2012;38(2):222-225.
doi:10.1016/j.joen.2011.10.017 .
Plotino, Gianluca, Costanzo, Alberto, Grande, Nicola M., Petrović, Renata, Testarelli, Luca, Gambarini, Gianluca, "Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments" in Journal of Endodontics, 38, no. 2 (2012):222-225,
https://doi.org/10.1016/j.joen.2011.10.017 . .
1
84
59
74