Clinical reproducibility of three electronic apex locators

Authorized Users Only
2011
Authors
Miletić, Vesna
Beljić-Ivanović, Katarina
Ivanović, Vladimir
Article (Published version)
Metadata
Show full item record
Abstract
Aim To compare the reproducibility of three electronic apex locators (EALs), Dentaport ZX, RomiApex A-15 and Raypex 5, under clinical conditions. Methodology Forty-eight root canals of incisors, canines and premolars with or without radiographically confirmed periapical lesions required root canal treatment in 42 patients. In each root canal, all three EALs were used to determine the working length (WL) that was defined as the zero reading and indicated by 'Apex', '0.0' or 'red square' markings on the EAL display. A new K-file of the same size was used for each measurement. The file length was fixed with a rubber stop and measured to an accuracy of 0.01 mm. Measurements were undertaken by two calibrated operators. Differences in zero readings between the three EALs in the same root canal were statistically analysed using paired t-tests with the Bonferroni correction, Bland-Altman plot and Linn's concordance correlation coefficients at alpha = 0.05. Results Mean and standard deviation v...alues measured by the three EALs showed no statistically significant differences. Identical readings by all three EALs were found in 10.4% of root canals. Forty-three per cent of readings differed by less than +/- 0.5 mm and 31.3% exceeded a difference of +/- 1 mm. Conclusions The clinical reproducibility of Dentaport ZX, RomiApex A-15 and Raypex 5 was confirmed with the majority of readings within the +/- 1.0 mm range. However, the small number of identical zero readings suggests that EALs are not reliable as the sole means of WL determination under clinical conditions.

Keywords:
anatomical foramen / apical constriction / clinical / electronic apex locators / reproducibility / working length determination
Source:
International Endodontic Journal, 2011, 44, 8, 769-776
Publisher:
  • Wiley-Blackwell, Malden

DOI: 10.1111/j.1365-2591.2011.01897.x

ISSN: 0143-2885

PubMed: 21599708

WoS: 000292336400008

Scopus: 2-s2.0-79960027301
[ Google Scholar ]
15
10
URI
https://smile.stomf.bg.ac.rs/handle/123456789/1620
Collections
  • Radovi istraživača
Institution/Community
Stomatološki fakultet
TY  - JOUR
AU  - Miletić, Vesna
AU  - Beljić-Ivanović, Katarina
AU  - Ivanović, Vladimir
PY  - 2011
UR  - https://smile.stomf.bg.ac.rs/handle/123456789/1620
AB  - Aim To compare the reproducibility of three electronic apex locators (EALs), Dentaport ZX, RomiApex A-15 and Raypex 5, under clinical conditions. Methodology Forty-eight root canals of incisors, canines and premolars with or without radiographically confirmed periapical lesions required root canal treatment in 42 patients. In each root canal, all three EALs were used to determine the working length (WL) that was defined as the zero reading and indicated by 'Apex', '0.0' or 'red square' markings on the EAL display. A new K-file of the same size was used for each measurement. The file length was fixed with a rubber stop and measured to an accuracy of 0.01 mm. Measurements were undertaken by two calibrated operators. Differences in zero readings between the three EALs in the same root canal were statistically analysed using paired t-tests with the Bonferroni correction, Bland-Altman plot and Linn's concordance correlation coefficients at alpha = 0.05. Results Mean and standard deviation values measured by the three EALs showed no statistically significant differences. Identical readings by all three EALs were found in 10.4% of root canals. Forty-three per cent of readings differed by less than +/- 0.5 mm and 31.3% exceeded a difference of +/- 1 mm. Conclusions The clinical reproducibility of Dentaport ZX, RomiApex A-15 and Raypex 5 was confirmed with the majority of readings within the +/- 1.0 mm range. However, the small number of identical zero readings suggests that EALs are not reliable as the sole means of WL determination under clinical conditions.
PB  - Wiley-Blackwell, Malden
T2  - International Endodontic Journal
T1  - Clinical reproducibility of three electronic apex locators
VL  - 44
IS  - 8
SP  - 769
EP  - 776
DO  - 10.1111/j.1365-2591.2011.01897.x
ER  - 
@article{
author = "Miletić, Vesna and Beljić-Ivanović, Katarina and Ivanović, Vladimir",
year = "2011",
abstract = "Aim To compare the reproducibility of three electronic apex locators (EALs), Dentaport ZX, RomiApex A-15 and Raypex 5, under clinical conditions. Methodology Forty-eight root canals of incisors, canines and premolars with or without radiographically confirmed periapical lesions required root canal treatment in 42 patients. In each root canal, all three EALs were used to determine the working length (WL) that was defined as the zero reading and indicated by 'Apex', '0.0' or 'red square' markings on the EAL display. A new K-file of the same size was used for each measurement. The file length was fixed with a rubber stop and measured to an accuracy of 0.01 mm. Measurements were undertaken by two calibrated operators. Differences in zero readings between the three EALs in the same root canal were statistically analysed using paired t-tests with the Bonferroni correction, Bland-Altman plot and Linn's concordance correlation coefficients at alpha = 0.05. Results Mean and standard deviation values measured by the three EALs showed no statistically significant differences. Identical readings by all three EALs were found in 10.4% of root canals. Forty-three per cent of readings differed by less than +/- 0.5 mm and 31.3% exceeded a difference of +/- 1 mm. Conclusions The clinical reproducibility of Dentaport ZX, RomiApex A-15 and Raypex 5 was confirmed with the majority of readings within the +/- 1.0 mm range. However, the small number of identical zero readings suggests that EALs are not reliable as the sole means of WL determination under clinical conditions.",
publisher = "Wiley-Blackwell, Malden",
journal = "International Endodontic Journal",
title = "Clinical reproducibility of three electronic apex locators",
volume = "44",
number = "8",
pages = "769-776",
doi = "10.1111/j.1365-2591.2011.01897.x"
}
Miletić, V., Beljić-Ivanović, K.,& Ivanović, V.. (2011). Clinical reproducibility of three electronic apex locators. in International Endodontic Journal
Wiley-Blackwell, Malden., 44(8), 769-776.
https://doi.org/10.1111/j.1365-2591.2011.01897.x
Miletić V, Beljić-Ivanović K, Ivanović V. Clinical reproducibility of three electronic apex locators. in International Endodontic Journal. 2011;44(8):769-776.
doi:10.1111/j.1365-2591.2011.01897.x .
Miletić, Vesna, Beljić-Ivanović, Katarina, Ivanović, Vladimir, "Clinical reproducibility of three electronic apex locators" in International Endodontic Journal, 44, no. 8 (2011):769-776,
https://doi.org/10.1111/j.1365-2591.2011.01897.x . .

DSpace software copyright © 2002-2015  DuraSpace
About Smile – School of dental Medicine dIgitaL archivE | Send Feedback

OpenAIRERCUB
 

 

All of DSpaceCommunitiesAuthorsTitlesSubjectsThis institutionAuthorsTitlesSubjects

Statistics

View Usage Statistics

DSpace software copyright © 2002-2015  DuraSpace
About Smile – School of dental Medicine dIgitaL archivE | Send Feedback

OpenAIRERCUB